www.ueitest.com • email: [email protected]om
DT304
Digital Temperature Logger
INSTRUCTION MANUAL
Vista de pagina 0
1 2 3 4 5 6 ... 19 20

Indice de contenidos

Pagina 1 - Digital Temperature Logger

1-800-547-5740 • Fax: (503) 643-6322www.ueitest.com • email: [email protected] Temperature LoggerINSTRUCTION MANUAL

Pagina 2 - TABLE OF CONTENTS

4. If you selected a user-defined logging interval (1 second to 23 hours and 59 minutes).• Press “ “ or “ “ until the display shows “h:m”

Pagina 3 - The DT304

Resetting the ThermometerIf you want to retore the thermometer settings as delivered from thefactory, press “ “ or “ “ until the display

Pagina 4 - Controls and Indicators

Holding the Displayed Temperature1. Press “ “ momentarily to freeze the readings on the display. The display shows “ HOLD “.2. Press “ “ m

Pagina 5

Using the Offset to Compensate for Probe ErrorsUse the offset option in “SETUP” to adjust temperature readings tocompensate for the errors of a specif

Pagina 6

Starting and Stopping LoggingMemory clear, and PC communications are in accessible during logging.Recall function is enabled during logging.1. Set the

Pagina 7 - Operating Instructions

5. Press “ “ momentarily or turn off the thermometer to stop viewing logged readings.NOTE: The thermometer calculates the minimum and maximum of

Pagina 8

These guidelines will help you attain long and reliable service from your meter:1. Calibrate your meter annually to ensure it meets original performa

Pagina 9

Battery ReplacementAlways use a fresh replacement battery of the specified size and type.Immediately remove the old or weak battery from the meter and

Pagina 10

16SpecificationsEnvironmentalGeneralOperating Temperature 14˚ to 122˚F (-10˚ to 50˚C)Storage Temperature -40˚ to 140˚F (-40˚ to 60˚C)Humidity Non cond

Pagina 11

TP6K K-Type Bead Thermocouple (a standard accessory)Electrical17Type K-Type, Chromel Alumel, bead styleTemperature range -40˚ to +900˚F (-40˚ to +480˚

Pagina 12

TABLE OF CONTENTSIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1Safety . . . . . . . . . . . . . . .

Pagina 13

DT304Digital Temperature LoggerLimited WarrantyThe DT304 is warranted to be free from defects in materials andworkmanship for a period of five years f

Pagina 14

IntroductionThe DT304Features include• IP67 Water/Dustproof rating• Large Backlit Multi paramater display• Convenient Probe Storage• Accepts J,K,T and

Pagina 15 - Maintenance

2Controls and Indicators1. Thermocouple T1, T2, T3 and T4 inputs2. Display3. Buttons4. USB jack213(top view)4

Pagina 16

LCD Display Functional Description1. Primary Display: T1/T2 or T3/T42. TYPE KJET: Type of thermocouple.3.SETUP: Setup is in progress.4. RECALL: The lo

Pagina 17

ButtonsNOTE: Press all the dual function buttons momentarily to activate theUpper Row Function ,and press these buttons for more than 1 secondto activ

Pagina 18 - Specifications

Press this button momentarily to recall or stop viewinglogged readings and MIN/MAX readings. Press this buttonfor more than 1 second to start or exit

Pagina 19

Setting the Date1. Entering Setup will show the first option of “ “2. Press “ENTER” and the display will show “ “3. Press “ “ or “

Pagina 20 - Limited Warranty

Changing the Auto-Power-Off Time (in minutes)The thermometer enters Auto-Power-Off mode if no button is pressedfor 30 minutes. Press “ “ to turn