KM9106 Operators Manual
1
1. ANALYSER LAYOUT AND FEATURES
1.1 Handset Features
Rubber boot
Remote lead socket
(15 pin D type)
Also used for
downloading data
to PC
4 line display
ON/OFF Key
UP key
DOWN key
ENTER/SETCAL
keys
Battery
Compartment
(back)
Commonly used keys
ON/OFF UP
Pages up through screens
Scrolls up through options, ie Fuel
ENTER & SET/CAL
Accepts parameters
Enters values
DOWN
Pages down through screens
Scrolls down through options
ON
OFF
SET/CAL
ENTER
Vista de pagina 0
1 2 3 4 5 6 ... 66 67

Indice de contenidos

Pagina 1 - Commonly used keys

KM9106 Operators Manual 11. ANALYSER LAYOUT AND FEATURES 1.1 Handset Features Rubber bootRemote lead socket(15 pin D type)Also used fordownloading

Pagina 2 - 1.2 Analyser Layout

KM9106 Operators Manual 10 4.3.2 Line Scroll Mode Line scroll mode allows you to customise the display. Use the key to change the bottom line o

Pagina 3

KM9106 Operators Manual 114.4 Sampling the Flue Gas Once the automatic calibration procedure has been completed and the specific fuel has been select

Pagina 4 - 1.4 Analyser Connections

KM9106 Operators Manual 12 4.5 Taking a Pressure Reading With the optional pressure module fitted a flue draught measurement can be made at any time.

Pagina 5 - 3. FIRST TIME USE

KM9106 Operators Manual 134.7 Normal Shutdown Sequence DO THIS EVERY TIME YOU USE THE ANALYSER. Remove the probe from the flue - TAKE CARE ! THE PRO

Pagina 6

KM9106 Operators Manual 14 5. USING THE KEYPAD 5.1 Basic Operation Basic operation of the keypad to change the display in PAGE and LINE SCROLL mode

Pagina 7 - 4. NORMAL START UP SEQUENCE

KM9106 Operators Manual 15UPPER PARAMETER KEYS ALWAYS Press UPPER FUNCTION first then SCALE to display Scaling setup paramete

Pagina 8

KM9106 Operators Manual 16 5.4 DISPLAY BACKLIGHT AND CONTRAST The display contrast can be adjusted to suit different lighting conditions and the bac

Pagina 9

KM9106 Operators Manual 176. USER SELECTABLE SETTING The following features are under your control at any time and can be changed as detailed later i

Pagina 10

KM9106 Operators Manual 18 they drift. This may happen if the sensor is taken to very high concentrations for long periods of time or over-ranged.

Pagina 11 - 4.4 Sampling the Flue Gas

KM9106 Operators Manual 19To Set The Time: TIME 16-01-12 SET As with the date each number in the time is to be set. Parameter Controls Op

Pagina 12 - KM9106 Operators Manual

KM9106 Operators Manual 2 1.2 Analyser Layout Connection leadfor OxygensensorOxygen Sensor Filter Bridgeor Sulphur FilterParticle FilterBatteryenclos

Pagina 13 - 4.9 HC104 Module

KM9106 Operators Manual 20 To Set A Standard Fuel: There are over 70 standard fuels programmed into Quintox. The fuels are arranged into tables for e

Pagina 14 - 5. USING THE KEYPAD

KM9106 Operators Manual 21To Set A User Fuel: If one of the standard fuels does not approximate to the one you are using in your boiler then it is po

Pagina 15

KM9106 Operators Manual 22 TO SET GROSS OR NETT EFFICIENCY Press the -- -- -- -- -- -- -- -- -- -- -- -- -- SET -- -- -- -- -- -- -- -- -- -- --

Pagina 16 - CONTRAST

KM9106 Operators Manual 23 Parameter Controls OptionsSELECT LANGUAGE Select using or key and ENGLISHSPANISHNETHERLAN

Pagina 17 - 6. USER SELECTABLE SETTING

KM9106 Operators Manual 24 Parameter Controls OptionsCOMPENSATION Select using or key and OFFON SET TEMPERATURE CELSIUS

Pagina 18

KM9106 Operators Manual 25Selecting YES allows you to set an oxygen reference value different to that shown in the fuel constants. For example 3% is s

Pagina 19

KM9106 Operators Manual 26 Parameter Controls OptionsREF NOx Select using or key and NOx = SUMNOx = NONOx = NO2 There are

Pagina 20

KM9106 Operators Manual 276.5 Inlet Temperature The Quintox uses as default the internal ambient sensor when calculating the Net temperature. If an

Pagina 21

KM9106 Operators Manual 28 6.6 Oxygen Calibration If used over long periods the Oxygen sensor may drift slightly and for maximum accuracy may requi

Pagina 22

KM9106 Operators Manual 29 CALIBRATE SENSOR Select using or key and YESNO Select YES SET ZERO ? NO Parameter Controls Opti

Pagina 23

KM9106 Operators Manual 31.3 Standard Probe Configuration

Pagina 24

KM9106 Operators Manual 30 6.8 CO Alarm It is possible to set a point in the range of the sensor so that it alarms and warns the user of a dangerou

Pagina 25

KM9106 Operators Manual 317. PRINTING INFORMATION While in any of the main displays a manual print can be obtained by pressing The display will show

Pagina 26

KM9106 Operators Manual 32 The remainder of this section explains the following :- Setting auto-timed printing or logging Allows information to be

Pagina 27

KM9106 Operators Manual 33The next screen enables or disables the integral printer. Default is ON. PRINTER ? ON Use or key to selec

Pagina 28

KM9106 Operators Manual 34 7.2 User Defined Printouts General Principle: A user defined printout can have a maximum of 40 lines. The contents of ea

Pagina 29

KM9106 Operators Manual 35The above example will send the Time and Oxygen reading to the printout every two minutes. To stop the Quintox Auto printin

Pagina 30

KM9106 Operators Manual 36 7.2.2 Master List of Printed Parameters The following list is a master list of printed parameters. Details of the measured

Pagina 31 - 7. PRINTING INFORMATION

KM9106 Operators Manual 378. QUINTOX LOGGING AND PC DOWNLOAD 8.1 Overview 8.1.1 Description Data is sent to the handset through the connection lead

Pagina 32

KM9106 Operators Manual 38 8.1.3 Logging Data Data logging is done with the handset operating in 2. QUINTOX CONTROL and by either pressing the

Pagina 33

KM9106 Operators Manual 398.3 Logger Control This facility allows access to all of the information stored in the handset; data can be cleared, viewed

Pagina 34

KM9106 Operators Manual 4 1.4 Analyser Connections

Pagina 35

KM9106 Operators Manual 40 LOG DATA ? NO ENTER YES to tell the logger to store data in memory and NO to disable the logging function. DOWNLO

Pagina 36

KM9106 Operators Manual 41ENTER the desired location to display the memory FROM. Entering numbers is detailed in section 6. LOG NO. DATE TIME BATTE

Pagina 37

KM9106 Operators Manual 42 8.4.1 Setting Up Your PC A standard communications package will be able to collect the data from the Quintox. Windows Term

Pagina 38 - 8.2 Batteries

KM9106 Operators Manual 438.4.2 Capturing Data From Quintox To capture data direct from the Quintox into a PC the RS232 port on the Quintox has to be

Pagina 39

KM9106 Operators Manual 44 9. CONTINUOUS MONITORING The Quintox is designed primarily as a portable emissions monitoring analyser. It can however be u

Pagina 40

KM9106 Operators Manual 4510. Main Purge The main purge facility for Quintox is used where longer sampling of flue gas is required. It should be used

Pagina 41

KM9106 Operators Manual 46 To access timed purge enter the following :- Enter the code 0000 at the following display :- ENTER SERVICE CODE --

Pagina 42 - 8.4.1 Setting Up Your PC

KM9106 Operators Manual 47Set YES or NO depending on whether auto zero is required or not. AUTO ZERO ? NO Note: During both Manual and Timed pur

Pagina 43

KM9106 Operators Manual 48 11. MAINTENANCE 11.1 Emptying and Cleaning the In-line Water Trap The water trap should be checked and emptied on a regula

Pagina 44 - 9. CONTINUOUS MONITORING

KM9106 Operators Manual 4911.3 Charging the Battery It is important that the battery is charged on a regular basis. The instrument constantly powers

Pagina 45 - 10. Main Purge

KM9106 Operators Manual 5 2. SAFETY WARNING This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases ar

Pagina 46

KM9106 Operators Manual 50 11.5 Changing the Printer Ribbon The printer ribbon cartridge will last for approximately two rolls of paper. Remove the p

Pagina 47

KM9106 Operators Manual 5112. PROBLEM SOLVING The following is a list of problems that may occur on the instrument through its operating life. If the

Pagina 48 - 11. MAINTENANCE

KM9106 Operators Manual 52 13. ANNUAL RE-CALIBRATION Whilst the sensors have an expected life of more than two years in normal use it is recommended t

Pagina 49

KM9106 Operators Manual 5314. HOW TO GET EXPERT HELP There will be occasions when despite having read the manual there will be problems that you can

Pagina 50

KM9106 Operators Manual 54 15. PRODUCT SPECIFICATION INSTRUMENT Parameter Resolution Accuracy Range Temp Measurement Flue Temperature Inlet Tempe

Pagina 51 - 12. PROBLEM SOLVING

KM9106 Operators Manual 55HANDSET Dimensions 220 mm long 120 mm high 50 mm wide Keypad 16 tactile keys with sounder Display 4 line LCD with back

Pagina 52 - 13. ANNUAL RE-CALIBRATION

KM9106 Operators Manual 56 AMBIENT OPERATING RANGE 0 - 40°C (+32°F to 104°F) 20 - 80% RH non condensing Storage: 0-50°C Maximum gas temperature a

Pagina 53 - 14. HOW TO GET EXPERT HELP

KM9106 Operators Manual 57APPENDICES A - Parameter Meanings The parameters and their meanings are detailed as follows : - DATE : Analyser date. See

Pagina 54 - 15. PRODUCT SPECIFICATION

KM9106 Operators Manual 58 NETT : Nett temperature calculated by deducting the internal AMBIENT temperature from the measured FLUE temperature. Di

Pagina 55

KM9106 Operators Manual 59CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by CO2. It gives an indication of the following :- •

Pagina 56

KM9106 Operators Manual 6 Printed header name and telephone number 7.1

Pagina 57 - APPENDICES

KM9106 Operators Manual 60 CxHy : Unburnt HydroCarbon reading in % of LEL of Methane, the sensor is calibrated with Methane. Displayed when a HydroCa

Pagina 58

KM9106 Operators Manual 61B. NOx CALCULATIONS ONLY an NO Sensor fitted. working in ppm: NOx referenced to NO The user can select the assumed NO

Pagina 59

KM9106 Operators Manual 62 NOx referenced to NO2 NOx in mg/m3 = NO in ppm multiplied by 2.05 multiplied by (1 + assumed NO2 percentage) or

Pagina 60

KM9106 Operators Manual 63ONLY an NO2 sensor fitted When there is only an NO2 sensor fitted the NOx function is disabled NO2 in mg/m3 = NO2 in ppm m

Pagina 61 - B. NOx CALCULATIONS

KM9106 Operators Manual 64 C. COMBUSTION EFFICIENCY CALCULATION The efficiency calculation is based upon British Standard BS845. This identifies thr

Pagina 62

KM9106 Operators Manual 65 Calculated data: Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unbu

Pagina 63 - ONLY an NO

KM9106 Operators Manual 66 D. CALCULATION OF FUEL DATA For any fuel not specified by Kane International the net calorific value, gross calorific valu

Pagina 64

KM9106 Operators Manual 67E. ELECTROMAGNETIC COMPATABILITY (CE) STATEMENT This product has been tested for compliance with the following generic

Pagina 65

KM9106 Operators Manual 74. NORMAL START UP SEQUENCE 4.1 Every Time You Use The Analyser BEFORE SWITCH-ON CHECK THAT: the Oxygen sensor is connec

Pagina 66 - D. CALCULATION OF FUEL DATA

KM9106 Operators Manual 8 And show the countdown screen :- -- -- -- -- -- -- -- -- -- -- -- -- -- 9 CALIBRATING 9 9 300 -- -- -- -- -- -- -- -- --

Pagina 67

KM9106 Operators Manual 94.3 Main Displays The main display can be changed to show the following :- • Page Mode displays 4 lines of data in set for